DOSE CALCULATION FOR THE HALF BLOCKED FIELDS DEFINED BY INDEPENDENT JAWS

Authors

  • Sivalee SURIYAPEE Department of Radiology Faculty of Medicine, Chulalongkorn University
  • Chotika JUMPANGERN Department of Radiology Faculty of Medicine, Chulalongkorn University
  • Chusak PUANGNAK Department of Radiology Faculty of Medicine, Chulalongkorn University

Keywords:

dose calculation, half blocked fields, independent jaws, asymmetric fields

Abstract

A system of dose calculation of a 6 MV photon beam from a Clinac 1800 linear acce -lerator for the half blocked fields defined by independent jaws was designed. The dose is computed as the product of the field size factors, the off axis ratios and the percentage depth dose which are determined from a data set of symmetric fields. The dosimetric measurements of these parameters for asymmetric fields were performed to evaluate the calculation method. The field size factors for asymmetric fields are within 1.1% compared to those for symmetric fields corrected for off axis ratios. Asymmetric field percent depth doses differ from those of symmetric field with maximum of 7.5% for field sizes range 4x4 to 20x20 cm. and depth down to 20 cm. The measured isodose curves of asymmetric fields show about 4% reduction of dose near the field edge closed to the flattening filter center. Calculation of the doses for half beam by this method have been checked by the measurement of the dosage at 1.5, 5 and 10 cm depth, the result shows an agreement within 1.3% which is acceptable.

Downloads

Download data is not yet available.

Metrics

Metrics Loading ...

References

Palta JR, Ayyangar KM, Suntharalingam N. Dosimetric characteristics of a 6 MVphoton beam from a linear accelerator with asymmetric col- limator jaws. Int J Radiation Oncology Biol Phys 1988; 14 (2): 383-387.

Slessinger ED, Gerber RL, Harms WB, et al. Independent collimator dosimetry for a dual photon energy linear accelerator. Int J Radiation Oncology Biol Phys 1993; 27 (3): 681-687.

Khan FM, Gerbi BJ, Deibel FC. Dosimetry of asymmetric x-ray collimators. Med Phys 1986; 13 (6):936-941.

Downloads

Published

2023-04-03

How to Cite

1.
SURIYAPEE S, JUMPANGERN C, PUANGNAK C. DOSE CALCULATION FOR THE HALF BLOCKED FIELDS DEFINED BY INDEPENDENT JAWS. ASEAN J Radiol [Internet]. 2023 Apr. 3 [cited 2024 Nov. 22];2(2):119-24. Available from: https://asean-journal-radiology.org/index.php/ajr/article/view/325

Issue

Section

Original Article

Most read articles by the same author(s)