SPATIAL RESOLUTION PERFORMANCE OF WHOLE-BODY LSO PET/CT SCANNER

Authors

  • Chatchai NAVIKHACHEEVIN Department of Radiology, King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Bangkok, Thailand

Keywords:

LSO, PET/CT, Biograph 16 HI-REZ

Abstract

Spatial resolution of Biograph 16 HI-REZ PET/CT system has been performed following the NEMA NU 2-2001 protocols. Three types of point source made with capillary tube, tiny foam bead, and ion exchanged resin were used for this purpose. ImageJ software was used to analyze the reconstruction image on PC. The results from three types of point source were in acceptance limit. In this study, point source made with resin was easy to be prepared and small size for spatial resolution performance.

Downloads

Download data is not yet available.

Metrics

Metrics Loading ...

References

Pasawang P, Sontrapompol T, Navikhacheevin C, Krisanachinda A. Performance Evaluation ofa Biograph 16 Hi-Rez PET Scanner at King Chulalongkom Memorial Hospital. Thai JRT 2006; 31: 9-14. 2.

Turkington TG. Introduction to PET Instrumenttation. J Nucl Med Technol 2001; 29: 1-8 3

Kennedy JA, Israel O, Frenkel A, Bar -Shalom, Azhari H. Super-Resolution in PET Imaging. IEEE Trans Med Imaging 2006; 25: 137-147

Martinez MJ, BercierY, Schwaiger M, Ziegler SI. PET/CT Biograph Sensation 16: Performance improvement using faster electronics. Nuklearmedizin 2006; 45: 126-133

National Electrical Manufactures Association. NEMA Standards Publication NU 2-2001: Performance Measurements of Positron Emission Tomographs, Rosslyn, VA: National Electrical Manufactures Association; 2001

Karp JS, Daube- Witherspoon ME, Hoffman E, et al. Performance Standard in Positron Emission Tomography. J Nucl Med 1991; 32: 2342-2350

Daube- Witherspoon ME, Karp JS, Casey ME.., et al. PET Performance Measurements Using the NEMA NU 2-2001 Standard. J Nucl Med 2002; 43: 1398-1409

Herzog H, Tellmann L, Hocke C, et al. NEMA NU 2-2001 Guided Performance Evaluation of Four Siemens ECAT PET Scanners. IEEE Trans Nucl Sci 2004; 51: 2662-2669

Brambilla M, Secco C, Dominietto M, et al. Performance Characteristics Obtained fora New 3-Dimensional Lutetium Oxyorthosilicate -Based Whole-Body PET/CT Scanner with the National Electrical Manufacturers Association NU 2-2001 Standard. J Nucl Med 2005; 46: 2083-2091

Bercier Y, Casey M, Young J, Wheelock J, Gremillion T. LSO PET/CT Pico performance improvements with ultra Hi-Rez option, IEEE Nuclear Science Symposium Conference Record, Rome, Institute of Electrical and Electronics Engineers, Inc. 2004; 7: 4038-4042

Erdi YE, Nehmeh SA, Mulnix T, Humm JL, Watson CC. PET Performance Measurements for an LSO-Based Combined PET/CT Scanner Using the National Electrical Manufacturers Association NU 2-2001 Standard. J Nucl Med 2004; 45: 813-821

CPS Innovation: Product Specification LSO PET/CT HI-REZ 16, Knoxville, TN: CPS Inc.: 2004