SPATIAL RESOLUTION PERFORMANCE OF WHOLE-BODY LSO PET/CT SCANNER
Keywords:
LSO, PET/CT, Biograph 16 HI-REZAbstract
Spatial resolution of Biograph 16 HI-REZ PET/CT system has been performed following the NEMA NU 2-2001 protocols. Three types of point source made with capillary tube, tiny foam bead, and ion exchanged resin were used for this purpose. ImageJ software was used to analyze the reconstruction image on PC. The results from three types of point source were in acceptance limit. In this study, point source made with resin was easy to be prepared and small size for spatial resolution performance.
Downloads
Metrics
References
Pasawang P, Sontrapompol T, Navikhacheevin C, Krisanachinda A. Performance Evaluation ofa Biograph 16 Hi-Rez PET Scanner at King Chulalongkom Memorial Hospital. Thai JRT 2006; 31: 9-14. 2.
Turkington TG. Introduction to PET Instrumenttation. J Nucl Med Technol 2001; 29: 1-8 3
Kennedy JA, Israel O, Frenkel A, Bar -Shalom, Azhari H. Super-Resolution in PET Imaging. IEEE Trans Med Imaging 2006; 25: 137-147
Martinez MJ, BercierY, Schwaiger M, Ziegler SI. PET/CT Biograph Sensation 16: Performance improvement using faster electronics. Nuklearmedizin 2006; 45: 126-133
National Electrical Manufactures Association. NEMA Standards Publication NU 2-2001: Performance Measurements of Positron Emission Tomographs, Rosslyn, VA: National Electrical Manufactures Association; 2001
Karp JS, Daube- Witherspoon ME, Hoffman E, et al. Performance Standard in Positron Emission Tomography. J Nucl Med 1991; 32: 2342-2350
Daube- Witherspoon ME, Karp JS, Casey ME.., et al. PET Performance Measurements Using the NEMA NU 2-2001 Standard. J Nucl Med 2002; 43: 1398-1409
Herzog H, Tellmann L, Hocke C, et al. NEMA NU 2-2001 Guided Performance Evaluation of Four Siemens ECAT PET Scanners. IEEE Trans Nucl Sci 2004; 51: 2662-2669
Brambilla M, Secco C, Dominietto M, et al. Performance Characteristics Obtained fora New 3-Dimensional Lutetium Oxyorthosilicate -Based Whole-Body PET/CT Scanner with the National Electrical Manufacturers Association NU 2-2001 Standard. J Nucl Med 2005; 46: 2083-2091
Bercier Y, Casey M, Young J, Wheelock J, Gremillion T. LSO PET/CT Pico performance improvements with ultra Hi-Rez option, IEEE Nuclear Science Symposium Conference Record, Rome, Institute of Electrical and Electronics Engineers, Inc. 2004; 7: 4038-4042
Erdi YE, Nehmeh SA, Mulnix T, Humm JL, Watson CC. PET Performance Measurements for an LSO-Based Combined PET/CT Scanner Using the National Electrical Manufacturers Association NU 2-2001 Standard. J Nucl Med 2004; 45: 813-821
CPS Innovation: Product Specification LSO PET/CT HI-REZ 16, Knoxville, TN: CPS Inc.: 2004
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2023 The ASEAN Journal of Radiology
This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
Disclosure Forms and Copyright Agreements
All authors listed on the manuscript must complete both the electronic copyright agreement. (in the case of acceptance)