RUPTURE OF RADIOPAQUE TIP OF TORCON NB CATHETER

Authors

  • Komgrit TANISARO Department of Radiology Hospital of Prince of Songkla University Hatyai

Keywords:

complication of angiography, catheter complication

Abstract

Radiopaque Tip Polyethylene diagnostic catheter was used when selecting the superior mesenteric artery before transarterial chemoembolization in hepatocellular carcinoma patient. A breaking of the catheter at the junction between the radiopaque tip and its shaft was noted. The broken tip was retrieved using 3-F vascular retreival forcep and surgical cutdown. It was claimed that opacification of the tip could reduce contrast consumption and fluoroscopy time. The use of the radiopaque tip “Torcon NB” catheter requires a high concern regarding possible catheter rupture.

Downloads

Download data is not yet available.

Metrics

Metrics Loading ...

References

Flisher RG, Ferreyro R. Evaluation of current techniques for non surgical removal of intravascular foreign bodies. AJR 1978; 130:541

Egglin TK, Dickey KW, Rosenblatt M et al., Retrieval of intravascular foreign bodies: experience in 32 cases. AJR 1995; 164:1259

Yang FS, Ohta I, Chiang HJ et al., Nonsurgical retrieval of intravascular foreign body: experience of 12 cases. EurJ Radiol 1994;18:1

Weiele JB, Sheline ME, Cope C. Expandable intravascular catheter: percutaneous use for endoliminal retrieval. Radiology 1992:185:604

Dondelinger RF, Lepoutre B, Kurdziel JC. Percutaneous vascular foreign body retrieval: expereince of an 11 years period. EurJ Radiol 1991;12:4

Ulflacker R, Lima S, Melichar AC. Intravascular foreign bodies: percutaenous retrieval. Radiology 1986;160:731

Baum S,Pentecost MJ. Abrams’ Angiography-Percutaneous vascular and nonvascular foreign body retrieval. Boston: Little, brown; 1997, pp. 959

Savader SJ, Trerotola SO. Venous Interven tional Radiology with clinical perspectivesforeign body retrieval. New York: Thieme; 1996, pp. 459

Paul L, Villar R, Sanchez A. Rupture of Radiopaque Tip of the Torcon MR Catheter (letter). JVIR 1997:8:300

Chomyn JJ, Graven WM. Groves BM et al., Percutaneous removal of a Gianturco coil from the pulmonary artery with use of intravascular forceps. JVIR 1991;2:105

Selby JB, Tegtmeyer CJ, Bittner GM. Experience with new retrieval forceps for foreign body removal in the vascular, urinary and biliary systems. Radiology 1990; 176:535

Downloads

Published

2023-04-10

How to Cite

1.
TANISARO K. RUPTURE OF RADIOPAQUE TIP OF TORCON NB CATHETER. ASEAN J Radiol [Internet]. 2023 Apr. 10 [cited 2024 Nov. 21];5(2):187-9. Available from: https://asean-journal-radiology.org/index.php/ajr/article/view/485

Issue

Section

Original Article

Similar Articles

<< < 1 2 3 4 5 6 7 > >> 

You may also start an advanced similarity search for this article.