DOSE CALCULATION FOR THE HALF BLOCKED FIELDS DEFINED BY INDEPENDENT JAWS
Keywords:
dose calculation, half blocked fields, independent jaws, asymmetric fieldsAbstract
A system of dose calculation of a 6 MV photon beam from a Clinac 1800 linear acce -lerator for the half blocked fields defined by independent jaws was designed. The dose is computed as the product of the field size factors, the off axis ratios and the percentage depth dose which are determined from a data set of symmetric fields. The dosimetric measurements of these parameters for asymmetric fields were performed to evaluate the calculation method. The field size factors for asymmetric fields are within 1.1% compared to those for symmetric fields corrected for off axis ratios. Asymmetric field percent depth doses differ from those of symmetric field with maximum of 7.5% for field sizes range 4x4 to 20x20 cm. and depth down to 20 cm. The measured isodose curves of asymmetric fields show about 4% reduction of dose near the field edge closed to the flattening filter center. Calculation of the doses for half beam by this method have been checked by the measurement of the dosage at 1.5, 5 and 10 cm depth, the result shows an agreement within 1.3% which is acceptable.
Downloads
Metrics
References
Palta JR, Ayyangar KM, Suntharalingam N. Dosimetric characteristics of a 6 MVphoton beam from a linear accelerator with asymmetric col- limator jaws. Int J Radiation Oncology Biol Phys 1988; 14 (2): 383-387.
Slessinger ED, Gerber RL, Harms WB, et al. Independent collimator dosimetry for a dual photon energy linear accelerator. Int J Radiation Oncology Biol Phys 1993; 27 (3): 681-687.
Khan FM, Gerbi BJ, Deibel FC. Dosimetry of asymmetric x-ray collimators. Med Phys 1986; 13 (6):936-941.
Downloads
Published
How to Cite
Issue
Section
License
Copyright (c) 2023 The ASEAN Journal of Radiology
This work is licensed under a Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International License.
Disclosure Forms and Copyright Agreements
All authors listed on the manuscript must complete both the electronic copyright agreement. (in the case of acceptance)